Abstract

Aortic valve replacement surgery technique has been revolutionised with the introduction of sutureless valves. Perceval sutureless aortic valve (PSV) is one of the most widely used examples. However, oversizing the PSV has been reported to be an issue. After a case of PSV stent recoil in our institution, a new technique was developed, where only the white obturator is used to size the valve. Recently, this technique was also mentioned by different authors. With this technique, we achieved low rates of paravalvular leaks and pacemaker implantation, with good transvalvular gradients. Therefore, we recommend this technique over the manufacturer’s instructions.

Item ID: 89987
Item Type: Article (Short Note)
ISSN: 2251-6638
Keywords: Perceval, Sutureless Aortic Valve, Technique
Copyright Information: This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 Unported License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.