Commissioning measurements on an Elekta Unity MR-Linac
Powers, Marcus, Baines, John, Crane, Robert, Fisher, Chantelle, Gibson, Stephen, Marsh, Linda, Oar, Bronwyn, Brodmann, Ariadne, Simpson-Page, Emily, Van der Walt, Marchant, and de Vine, Glenn (2022) Commissioning measurements on an Elekta Unity MR-Linac. Physical and Engineering Sciences in Medicine, 45. pp. 457-473.
|
PDF (Publisher Accepted Version)
- Published Version
Available under License Creative Commons Attribution. Download (1MB) | Preview |
Abstract
Magnetic resonance-guided radiotherapy technology is relatively new and commissioning publications, quality assurance (QA) protocols and commercial products are limited. This work provides guidance for implementation measurements that may be performed on the Elekta Unity MR-Linac (Elekta, Stockholm, Sweden). Adaptations of vendor supplied phantoms facilitated determination of gantry angle accuracy and linac isocentre, whereas in-house developed phantoms were used for end-to-end testing and anterior coil attenuation measurements. Third-party devices were used for measuring beam quality, reference dosimetry and during treatment plan commissioning; however, due to several challenges, variations on standard techniques were required. Gantry angle accuracy was within 0.1°, confirmed with pixel intensity profiles, and MV isocentre diameter was < 0.5 mm. Anterior coil attenuation was approximately 0.6%. Beam quality as determined by TPR20,10 was 0.705 ± 0.001, in agreement with treatment planning system (TPS) calculations, and gamma comparison against the TPS for a 22.0 × 22.0 cm2 field was above 95.0% (2.0%, 2.0 mm). Machine output was 1.000 ± 0.002 Gy per 100 MU, depth 5.0 cm. During treatment plan commissioning, sub-standard results indicated issues with machine behaviour. Once rectified, gamma comparisons were above 95.0% (2.0%, 2.0 mm). Centres which may not have access to specialized equipment can use in-house developed phantoms, or adapt those supplied by the vendor, to perform commissioning work and confirm operation of the MRL within published tolerances. The plan QA techniques used in this work can highlight issues with machine behaviour when appropriate gamma criteria are set.
Item ID: | 71517 |
---|---|
Item Type: | Article (Research - C1) |
ISSN: | 2662-4737 |
Keywords: | MR-Linac, Commissioning, Elekta Unity |
Related URLs: | |
Copyright Information: | © Crown 2022. This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. |
Date Deposited: | 20 May 2022 05:49 |
FoR Codes: | 51 PHYSICAL SCIENCES > 5105 Medical and biological physics > 510502 Medical physics @ 100% |
SEO Codes: | 28 EXPANDING KNOWLEDGE > 2801 Expanding knowledge > 280120 Expanding knowledge in the physical sciences @ 25% 20 HEALTH > 2002 Evaluation of health and support services > 200206 Health system performance (incl. effectiveness of programs) @ 75% |
Downloads: |
Total: 636 Last 12 Months: 7 |
More Statistics |