Sources of out-of-field dose in MRgRT: an inter-comparison of measured and Monaco treatment planning system doses for the Elekta Unity MR-linac

Baines, John, Powers, Marcus, and Newman, Glen (2021) Sources of out-of-field dose in MRgRT: an inter-comparison of measured and Monaco treatment planning system doses for the Elekta Unity MR-linac. Physical and Engineering Sciences in Medicine, 44. pp. 1049-1059.

[img] PDF (Publisher Accepted Version) - Published Version
Restricted to Repository staff only

View at Publisher Website: https://doi.org/10.1007/s13246-021-01039...
 
3
2


Abstract

With the clinical introduction of MR-linacs, out-of-field dose (OFD) associated with head leakage/scatter (HLS), spiralling contaminant electrons (SCE) and the electron streaming effect (ESE) is of interest. To investigate HLS and SCE, EBT3 film on solid water 5.0 cm beyond each edge of a 10.0 × 10.0 cm2 field was used to determine depth-dose for 0 T and 1.5 T, in the isocentric plane. Additionally, ESE induced by the anterior imaging coil was quantified and the experimental arrangements to measure SCE and ESE were modelled using Monaco. For a clinical treatment of supraclavicular nodal disease, Monaco OFD was compared to in vivo measurements. For 0 T, depth-dose was isotropic and surface dose was approximately 4.4% of Dmax. With 1.5 T surface doses were approximately 3.8% of Dmax at ± Y (IEC61217), compared to 2.6% and 0.6% of Dmax at − X and X, respectively. For both field strengths, the TPS depth-dose variation was consistent with experimental trends; however, near surface doses calculated at ± Y differed significantly from measurements. For the field sizes investigated, measured coil ESE dose was between 9.0 and 28.0% of Dmax and Monaco coil ESE was less than measured by up to 13.0%. OFD in 0 T and 1.5 T are comparable at ± Y, inconsistent with previous work. Anterior coil ESE should be mitigated during treatment and for the clinical case investigated, in vivo OFD was within 2σ of TPS calculations. Monaco overestimates near surface SCE and underestimates coil ESE.

Item ID: 71514
Item Type: Article (Research - C1)
ISSN: 2662-4737
Keywords: MRgRT, Spiralling contaminant electrons, Electron streaming effect, Monaco unity
Related URLs:
Copyright Information: © Crown 2021
Research Data: https://doi.org/10.25903/snzx-c059
Date Deposited: 03 Feb 2022 22:54
FoR Codes: 51 PHYSICAL SCIENCES > 5105 Medical and biological physics > 510502 Medical physics @ 100%
SEO Codes: 28 EXPANDING KNOWLEDGE > 2801 Expanding knowledge > 280120 Expanding knowledge in the physical sciences @ 50%
20 HEALTH > 2002 Evaluation of health and support services > 200206 Health system performance (incl. effectiveness of programs) @ 50%
Downloads: Total: 2
More Statistics

Actions (Repository Staff Only)

Item Control Page Item Control Page